Technical Specification and Data

 
Measurement Range Carbon Dioxide 0-255 mmHg (0-25.5 kPA). Oxygen 0-800 mmHg (0-99.9 kPA)
CO2 Metabolic Correction Displayed pCO2 is compensated for metabolic factors related to the temperature effect and CO2 production.
Parameter Alerts CO2 and O2 limit alerts, audible and visible. User selectable, continuously displayed, retained in memory.
Alert Silence Two minute alert silence with LED indicator. Audio Off with LED indicator (user can disable if needed).
Temperature Section User selectable 37-45° C in 0.5° C increments. Can be continuously displayed. Audible & visible alarm if actual/set temperatures differ by >0.2° C. Two fully independent thermistors monitor and control temperature. Heater power shutdown with audible & visible alarm at over-temp (45.5° C) or if thermistors differ by >0.5°.
Site Timer Automatically tracks sensor site duration, provides audible and visible alarms, automatic shutdown of heater power once timer expires. User selectable durations in 0.5 hr increments up to 5, 8, or 12 hour maximum, plus off.
Local Power The power required to maintain sensor temperature is measured and displayed. Range: 0-800 mW. User selectable power reference point allows for quick power trend analysis (LPR).
Real Time Trends Continuously displayed user selectable/scalable real time 10, 20, 30 minute trends of CO2, O2 or Local Power.
Trend/Histogram Memory Battery backed 24 hour trends, 8 second resolution. Trend/Histogram displays 30 min, 2, 4, 8, 12 hours.
External Devices Standard RS232 serial outputs: NovaCARD, computer, printer, others.
Optional Analog Module, CO2, O2 and LP.
System
  • Power: 100-120/200-240 VAC, 50-60 Hz, 40 VA.
     
  • Battery: Greater than 3 hour operation, 12 hour maximum recharge, lead-acid gel-cell.
     
  • Display Type: Cold Cathode Display (CCD), 5 × 1.5 inches, W × H (12.7 × 3.8 cm). User adjustable contrast.
     
  • Monitor Size: 3.3 × 9.0 × 8.0 inches, H × W × D (8.4 × 22.8 × 20 cm). Weight: 8 pounds (3.6 kg).
     
  • Temperature: Operating 10-40° C (50-104° F), Storage -10-55° C (14-131° F). Relative humidity: 0-90%.
     
  • Altitude/Pressure: Automatic barometric pressure compensation. Range: 560-800 mmHg (71.9-106.6 kPa).
     
  • Safety/Regulatory: Designed to comply with EC and TUV (IEC 601-1), UL (2601), and CSA (601-1-M90) standards.

 